Safety Notice: Posted or Updated on 23 Jun 2023

Abbott has recently discovered that when using smartphones with the Android 13 Operating System (OS), customers may experience extended periods of signal loss in the FreeStyle LibreLink and FreeStyle Libre 3 Android apps.

During these extended periods of signal loss, users of glucose alarms will not receive low glucose or high glucose alarms and FreeStyle Libre 3 users will not receive glucose readings.

Please click the link below to access the Safety Notice:

Urgent Field Safety Notice